Description
The IC3 Ion Chamber is a battery operated, auto ranging, portable ion chamber survey meter designed for highly stable and accurate measurement of dose rates and integrated dose of gamma, x-ray and beta radiation.
New – a sealed Electrometer minimizes noise and maximizes stability and accuracy
New – a double layered Mylar cover to improve Beta responses and capability to accuratelymeasure low energy radiation.
The IC3 Ion Chamber is ideal for use in nuclear power plants, nuclear medicine, radiography and radiotherapy facilities, life science laboratories, nuclear research centers and in other industrial applications.
The IC3 Ion Chamber is capable of measuring X-Ray bursts from 50 nanosecond pulses which makes it an ideal safety tool for X-Ray and NDT applications.
TECHNICAL INFORMATION
- Measuring Range: 1 µSv/h to 1 Sv/h (0.1 mR/h to 100 R/h)
- Accuracy (Cs-137): ±10% of reading within measuring range
- Sensitivity: From 0.1 mR/h to 250 mR/h: 172 cps/mR/h | From 250 mR/h to 50 R/h: 1.72 cps/mR/h
- Gamma Energy Dependence: Better than ±20% from 20 keV to 1.3MeV (related to 137Cs)
- Beta Energy Dependence: Better than ±20% from 200 keV
- Angular Dependence (Cs-137): Less than ±30% (for ±90 deg. front direction)
- Ion Chamber Volume: 350 cc